Syringeless CT power injectors (also called pressure injectors) have been developed as a solution to reduce contrast fluid waste. The Joint Commission does not allow the reuse of unused cans from disposable syringes, so the syringeless option allows facilities to use contrast media as efficiently as possible. A syringeless CT (computed tomography) contrast media injector permits delivery to be made directly from standard contrast media bottles, featuring multi-dosing and programmable automatic saline flush. This higher loading capacity for contrast agent ( up to two contrast bottles or 1000 ml) and a bag of saline at initial setup, enables to perform multiple contrast enhanced CT scans without reloading contrast agent; resulting in increased efficiency and less waste. See also Single-Head Contrast Media Injector, Dual-Head CT Power Injector, CT Power Injector, Multi-Head Contrast Media Injector. Further Reading: News & More:
•
Contrast media injectors are part of the medical equipment used to deliver fluids in examinations such as CT, MRI, fluoroscopy and angiography. Many of these diagnostic imaging procedures include the administration of intravenous contrast agents to enhance the blood and perfusion in tissues. Mainly there are two types of injector technology: •
Piston-based systems use a plunger/piston to move a piston in the cylinder of a reservoir, which works in two directions
to first fill the reservoir and then deliver the fluid from the reservoir to the patient, similar to a hand-held syringe.
•
Peristaltic-pump-based systems operate as rotary pumps that use rollers to compress sections of flexible tubing, drawing fluid directly
from the supply source and delivering it to the patient.
See also Single-Head Contrast Media Injector, Dual-Head CT Power Injector, Syringeless CT Power Injector. The use of x-ray contrast agents in computed tomography (CT) began with a hand injection by the radiologist in the scan room. During its history, CT scanners have made great improvements in speed and image quality. Actual CT systems with multiple detectors allow scan times of a few seconds per body region. Some CT protocols require multiphase scans, where a body region is imaged with a single bolus of contrast in different blood flow phases. Automatic power (pressure) contrast media injectors are required to provide precise control of flow rate, volume and timing of injection. The use of a saline bolus following contrast administration reduces the volume of contrast required. Most relevant topics for the use of a power injector in medical imaging procedures such as contrast enhanced computed tomography (CECT): •
Avoidance of microbiologic contamination;
•
workflow efficiency in the use of the contrast media injector;
•
contrast cost and waste volume;
•
reimbursement.
Must have basic injector control options: •
Flow rate with a usual range from 0.1 to 10 mL/sec in 0.1 mL/sec increments; some injectors can be set to inject in ml/min or
ml/hour;
•
pressure limit typically programmable from 50 psi to 300 psi in 1 psi increments
(also displayable in kPa and kg/cm²).
Examples of other injector control options: •
•
pre-filled syringes; the compatibility with many selected syringes makes it easy to change and select the appropriate contrast medium for each patient;