Contrast Media Injector
Contrast media injectors are part of the medical equipment used to deliver fluids in examinations such as CT, MRI, fluoroscopy and angiography. Many of these diagnostic imaging procedures include the administration of intravenous contrast agents to enhance the blood and perfusion in tissues.

Mainly there are two types of injector technology:
Piston-based systems use a plunger/piston to move a piston in the cylinder of a reservoir, which works in two directions to first fill the reservoir and then deliver the fluid from the reservoir to the patient, similar to a hand-held syringe.
Peristaltic-pump-based systems operate as rotary pumps that use rollers to compress sections of flexible tubing, drawing fluid directly from the supply source and delivering it to the patient.

See also Single-Head Contrast Media Injector, Dual-Head CT Power Injector, Syringeless CT Power Injector.

The use of x-ray contrast agents in computed tomography (CT) began with a hand injection by the radiologist in the scan room. During its history, CT scanners have made great improvements in speed and image quality. Actual CT systems with multiple detectors allow scan times of a few seconds per body region. Some CT protocols require multiphase scans, where a body region is imaged with a single bolus of contrast in different blood flow phases. Automatic power (pressure) contrast media injectors are required to provide precise control of flow rate, volume and timing of injection. The use of a saline bolus following contrast administration reduces the volume of contrast required.

Most relevant topics for the use of a power injector in medical imaging procedures such as contrast enhanced computed tomography (CECT):
Avoidance of microbiologic contamination;
workflow efficiency in the use of the contrast media injector;
contrast cost and waste volume;
reimbursement.

Must have basic injector control options:
Flow rate with a usual range from 0.1 to 10 mL/sec in 0.1 mL/sec increments; some injectors can be set to inject in ml/min or ml/hour;
volume range from 1 mL to 200 mL for contrast and saline phases;
pressure limit typically programmable from 50 psi to 300 psi in 1 psi increments (also displayable in kPa and kg/cm²).

Examples of other injector control options:
Warmer/heater; an increase in temperature of the contrast medium results in a decrease in its viscosity; warmed contrast media are less viscous and offer lesser resistance;
pre-filled syringes; the compatibility with many selected syringes makes it easy to change and select the appropriate contrast medium for each patient;
injection reports accessible via RIS/PACS for dose management systems and records of prior injections.

Osteo CT
An Osteo CT or quantitative computed tomography (QCT) is used to measure bone mineral density (BMD). The high contrast discrimination of computed tomography can be used to examine the central skeleton for osteoporosis. Common CT scanners require a standard of reference to properly perform quantitative tissue analysis.
Osteo CT is the most accurate bone densitometry study, but is not widely available and delivers more radiation to the patient than dual energy x-ray absorptiometry.
Tomography
Tomography is imaging by sections or sectioning to obtain images of slices through objects like the human body. Tomography is derived from the Greek words 'to cut or section' (tomos) and 'to write' (graphein). A device used in tomography is called a tomograph, while the image produced is a tomogram.
The first medical applications utilized x-rays for images of tissues based on their x-ray attenuation coefficient. The mathematical basis for tomographic imaging was laid down by Johann Radon. This type of imaging is used in different medical applications as for example computed tomography, ultrasound imaging, positron emission tomography and magnetic resonance imaging (MRI) also called magnetic resonance tomography (MRT).
Conventional x-ray tomographic techniques show organ structures lying in a predetermined plane (the focal plane), while blurring the tissue structures in planes above and below by linear or complex geometrical motion of the x-ray tube and film cassette.
Basically, computed tomography is the reconstruction of an image from its projections. In the strict sense of the word, a projection at a given angle is the integral of the image in the direction specified by that angle. The CT images (slices) are created in the axial plane, while coronal and sagittal images can be rendered by computer reconstruction.

See also Zonography, Computed or Computerized Axial Tomography, Resolution Element, Radiographic Noise, Intravenous Pyelogram.
Virtual Colonoscopy
Virtual colonoscopy provides a less invasive option to conventional polyp detection in the large intestine (colon and rectum). A virtual colonoscopy is a synthesis of a computed tomography (CT) scan, digital processing and virtual reality computer software.
A virtual colonoscopy is less invasive and more comfortable for patients than either conventional colonoscopy or a barium enema. No sedation is required and the examination takes less than 30 minutes.
A CT colonography offers a new option for a total colon evaluation and cancer detection and has the potential to be used for screening. A problem is the amount of information captured in a CT exam. Reviewing these images can be time-consuming and challenging.

See also Computed Tomography Enterography.
Barium Sulfate
Barium sulfate (BaSO4) is an inert and insoluble white powder with high density. Barium belongs chemically to the group of heavy metals. Mixed with water and additional ingredients (e.g., sweetening agents), barium sulfate is the preferred positive contrast agent for abdominal x-ray and computed tomography examinations. The extremely low solubility of barium sulfate protects patients from absorbing harmful amounts of the metal (water soluble metal compounds are often highly toxic). The high density in x-ray examinations is related to the high atomic number, since large nuclei absorb x-rays much better than smaller nuclei.
Barium sulfate agents for opacification of the gastrointestinal tract are not absorbed or metabolized and are resistant to dilution. These contrast agents are opaque white suspensions and usually swallowed or administered as an enema. They provide better delineation of mucosal details and are less expensive than water-soluble iodinated contrast media. The elimination rate is a function of gastrointestinal transit time. After GI application, it leaves the body with the feces.

Contraindications of barium sulfate products in case of known or suspected:
obstruction of the colon;
gastrointestinal tract perforation;
tracheoesophageal fistula;
obstructing lesions of the small intestine;
inflammation or neoplastic lesions of the rectum;
hypersensitivity to barium sulfate formulations;
recent rectal biopsy;
pyloric stenosis.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]